Pharmaceutics · Microbiology
Serial Dilution Calculator
Plan simple, serial and stepwise dilutions of a drug stock — volumes to pipette, diluent, every tube's concentration and a printable lab record.
Preparation
Printed on the lab result card and used to name a saved calculation.
Try a worked example
How the factor is entered
Dilution factor is X in 1:X — 1 part of solution in X parts total. 10 means 1:10, so each tube is C_new = C_start × 1/10 (× 0.1). The textbook form C_final = C_initial × dilution factor uses the fraction 1/X. X must be greater than 1.
Serial dilution
The same dilution factor in every tube.
Checked against the stock the first tube needs.
A whole number, 1–20.
Enter X of 1:X — e.g. 10 for 1:10 (× 0.1).
Reports the tube that reaches it. Your inputs are never changed.
Laboratory calculation card
Enter the stock concentration, number of steps, dilution factor and the volume in each tube.
C₁V₁ = C₂V₂ → V₁ = C₂ × V₂ ÷ C₁; diluent = V₂ − V₁
C_new = C_start × 1/X (dilution factor written 1:X)
Volume taken = Final volume ÷ X; diluent = Final volume − Volume taken
Overall factor = X₁ × X₂ × … × Xₙ; C_final = C₁ ÷ overall factor
Keep full volume. If every tube must still hold its final volume after the next tube has taken its share, each tube is made up to V + (volume the next tube takes). That depends on the next tube, so the volumes are worked out backwards from the last tube: made-up volumeᵢ = Vᵢ + made-up volumeᵢ₊₁ ÷ Xᵢ₊₁.
Worked example. Amoxicillin 10 mg/mL, 1:10 × 5 tubes of 10 mL: each tube takes 10 mL ÷ 10 = 1.00 mL and 9.00 mL of diluent. The concentrations are 1, 0.1, 0.01, 0.001 mg/mL and 0.1 µg/mL; the overall factor is 10⁵ = 100,000. With “keep full volume” on, T1 is made up to 11.1 mL (1.11 mL of stock + 10.0 mL diluent) so that it still holds 10.0 mL after T2 takes its 1.11 mL.
Worked example: doubling dilutions of an antibiotic for an MIC practical
For a broth-dilution practical you have an ampicillin stock of 1000 µg/mL and need five doubling (1:2) dilutions, each tube made up to 4 mL. What goes into each tube, and what concentration does the last tube hold? (Illustrative values.)
- Mode
- Serial dilution
- Stock concentration, C₁
- 1000 µg/mL
- Number of steps
- 5
- Dilution factor per step
- 2 (1:2)
- Final volume in each tube
- 4 mL
- Keep full volume
- Off
Volume taken per tube
Volume taken = final volume ÷ X = 4 mL ÷ 2 = 2.00 mL
Diluent per tube
Diluent = 4 mL − 2.00 mL = 2.00 mL
Concentration tube by tube
T1 = 1000 × 1/2 = 500 µg/mL T2 = 250, T3 = 125, T4 = 62.5 µg/mL T5 = 62.5 × 1/2 = 31.25 µg/mL
Overall factor and final concentration
Overall factor = 2⁵ = 32 C_final = 1000 ÷ 32 = 31.25 µg/mL
Materials
Stock required = T1 volume taken = 2.00 mL Total diluent = 5 × 2.00 mL = 10.0 mL
Put 2.00 mL of diluent in each of five tubes, transfer 2.00 mL along the series, and T5 holds 31.25 µg/mL (overall 1:32).
The calculator's summary reads: stock required 2.00 mL, total diluent 10.0 mL, overall dilution factor 1:32, final concentration 31.25 µg/mL. Because each tube gives 2.00 mL to the next, T1–T4 are left with 2.00 mL and T5 with 4.00 mL; switch on "keep full volume" if every tube must end at 4 mL.
Common mistakes
Assumptions and limitations
Keep learning
References
- Stockton SJ. Stoklosa and Ansel's Pharmaceutical Calculations. 16th ed. Philadelphia: Wolters Kluwer; 2021.
- Sinko PJ, editor. Martin's Physical Pharmacy and Pharmaceutical Sciences. 8th ed. Philadelphia: Wolters Kluwer; 2023.
Common questions
Yes, in this calculator. 1:10 means 1 part of solution in 10 parts total (1 mL stock + 9 mL diluent), so the concentration becomes one tenth. Some older texts write 1:9 for 1 part + 9 parts — check which convention your manual uses before you pipette.
When the single-step stock volume would be too small to pipette accurately (below about 10 µL on a typical micropipette), or when you need every intermediate concentration — for example the doubling dilutions of an MIC plate or the points of a standard curve.
Because nothing is taken out of it. Every other tube gives up a transfer to the next tube. Turn on "keep full volume" if every tube must end with the same volume — the calculator then makes each tube up a little larger.
Not without the molecular weight. Mass concentrations (mg/mL, g/L, µg/mL, % w/v) convert among themselves, and molar ones (M, mM, µM) convert among themselves. To cross over, first convert with mM = (mg/mL ÷ MW in g/mol) × 1000.
Pipetting errors compound. If each step is within 1 %, six steps can be about 6 % off in the last tube. Use calibrated pipettes, a fresh tip for each transfer, and mix every tube before taking from it.
Prepared by the PharmaWallah student team. Verify results with your textbook or instructor. Report an error · Editorial policy