Physiology Lab
RBC Count Calculator
Red cell count from an improved Neubauer hemocytometer — cells counted, squares, depth and dilution, worked step by step.
Sample
Printed on the lab result card.
Count and dilution
The total cells in all the squares you counted, and the dilution of the blood.
Quick setup
Total across every square counted.
1:200 means 1 volume of blood in 200 volumes of suspension.
Common dilutions
Counting squares
Which squares you counted, and the depth of your chamber. Check both against your manual.
Standard RBC count: 5 medium squares (4 corners + centre).
Improved Neubauer: 0.1 mm.
Live chamber readout
Area counted = n × a = 5 × 0.04 mm² = 0.2 mm²
Volume = 0.2 mm² × 0.1 mm = 0.02 mm³
Reference range (optional, educational)
No range is built in. Enter the one from your manual to add a comparison to the lab card.
Laboratory calculation card
Enter the cells counted and the dilution factor. The squares and depth start at the standard RBC setup.
Area counted (mm²) = number of squares × area of one square
Volume counted (mm³) = area counted × chamber depth
RBC (cells/mm³) = cells counted × dilution factor ÷ volume counted
Standard RBC count: 5 medium squares × 0.04 mm² × 0.1 mm = 0.02 mm³. The calculation factor is 200 ÷ 0.02 = 10,000, so 480 cells counted gives 480 × 10,000 = 4,800,000 cells/mm³.
1 mm³ = 1 µL; 4,800,000/µL = 4.80 × 10⁶/mm³ = 4.80 × 10¹²/L
Counting 80 smallest squares is the same area: 80 × 0.0025 mm² = 0.2 mm². A manual that dilutes 1:100 instead halves the factor to 100 ÷ 0.02 = 5,000. Manual factor mode skips the area and depth: count = cells counted × factor.
Common questions
Count cells touching the top and left boundary lines of each square and ignore those touching the bottom and right lines (or the other way round — just be consistent). Otherwise a cell lying on a shared line is counted in two neighbouring squares.
There are several million red cells per mm³ but only a few thousand white cells. Undiluted enough, red cells overlap and cannot be counted, so they need a ten-fold higher dilution and a much smaller counting area (0.2 mm² rather than 4 mm²).
Red cells swell and burst in a hypotonic fluid and shrink in a hypertonic one, which would lose cells or change how they look. Hayem's fluid is isotonic, so the cells keep their shape; its sodium sulphate also discourages rouleaux, and mercuric chloride acts as a preservative.
No. 80 smallest squares (5 groups of 16) are exactly the same 0.2 mm² as 5 medium squares, so the factor is unchanged. A 1:100 dilution halves the factor to 5,000. Use the quick-setup buttons, or Manual factor mode with your manual's factor.
Cells never spread perfectly evenly when the chamber fills. Sampling squares spread across the central large square averages out that unevenness. If the counts in the five squares differ widely, the chamber was probably filled badly and should be recharged.
Prepared by the PharmaWallah student team. Verify results with your textbook or instructor. Report an error · Editorial policy