Dialysis Membrane / Diffusion Calculator
The practical sheet's method, step by step: absorbance → C2 → C2/C1 → volume factor → B → ln(1 − B), then the graph against time, its slope and the rate constant.
Experiment
Printed on the lab record.
Calibration curve
Regression equation Y = a + bX (Y = absorbance, X = concentration)
Y = a + bX(Y = absorbance, X = conc. in µg/mL)
Concentration and volumes
C1 and C2 must be in the same unit — C2/C1 is unitless.
The practical sheet uses 5 mL.
The practical sheet uses 1500 mL.
Volume factor = 1 + V2/V1 = —
Time table
Time | Absorbance — add a row for every sample withdrawn
| # | Time(min) | Time(s) | Absorbance (Y)(AU) | Remove |
|---|---|---|---|---|
| Row 1 | 0 | |||
| Row 2 | 300 | |||
| Row 3 | 600 | |||
| Row 4 | 900 | |||
| Row 5 | 1,200 | |||
| Row 6 | 1,500 | |||
| Row 7 | 1,800 |
2–30 rows. Times must increase from row to row; rows are never re-ordered for you.
Load an example data set (you can edit every value)
Practical interpretation
How the slope of ln(1 − B) against time is reported
Laboratory calculation card
Enter the calibration line, C1, V1, V2 and every row's time and absorbance.
Time (s) = Time (min) × 60
C2 = (Y − a) / b (from Y = a + bX)
Volume factor = 1 + V2/V1
B = (C2/C1) × [1 + V2/V1]
ln value = ln(1 − B) — defined only when 1 − B > 0
ln(1 − B) = m·t + c, m = [nΣxy − (Σx)(Σy)] / [nΣx² − (Σx)²], c = [Σy − mΣx] / n
k = −slope (when the practical interprets the negative slope as the rate constant)
With V1 = 5 mL and V2 = 1500 mL the volume factor is 1 + 1500/5 = 301. Slope b belongs to the calibration line; B is the volume-corrected ratio. R² = 1 − SSE/SST for the ln(1 − B) line.
Keep learning
References
- Sinko PJ, editor. Martin's Physical Pharmacy and Pharmaceutical Sciences. 8th ed. Philadelphia: Wolters Kluwer; 2023.
- Harris DC, Lucy CA. Quantitative Chemical Analysis. 10th ed. New York: W. H. Freeman (Macmillan Learning); 2019.
Common questions
B has reached or passed 1, so 1 − B is zero or negative and has no logarithm. Check the absorbance, C1 and the volumes. That row is left out of the graph and the regression, and the calculator says so.
No. Slope b is the calibration line's slope (absorbance per unit concentration). B is the practical's volume-corrected ratio, (C2/C1) × [1 + V2/V1]. They only share a letter.
Either — the sheet records both. The slope and k change unit with the axis (k in s⁻¹ is k in min⁻¹ divided by 60); the intercept and R² do not change.
Choose “Report slope only”. The slope, intercept and R² are still calculated, and no rate constant is shown anywhere.
Prepared by the PharmaWallah student team. Verify results with your textbook or instructor. Report an error · Editorial policy